Posts

Microphone Acoustic Overload Point (AOP)

This sequence measures the acoustic overload point (AOP) of a microphone – the SPL required to produce 10% THD @ 1kHz from the microphone’s output.
A 1 kHz amplitude sweep from a calibrated source speaker is applied across a range from 120 dB SPL to 135 dB SPL.The recorded time waveform is then analyzed using the HarmonicTrak algorithm to produce a Fundamental and THD vs. Level curve. An intersection post processing function identifies where the THD curve intersects 10% and that level is is used in a limit step to produce a Pass/Fail AOP verdict.

More